Specifications
General |
Maximum voltage between any terminal and earth ground |
600 V |
Surge protection |
6 kV peak per IEC 61010-1 600 V CAT III, Pollution Degree 2 |
Display |
Digital: 6,000 counts, updates 4/sec |
Bar graph |
33 segments, updates 32/sec |
Operating temperature |
-10 °C to + 50 °C |
Storage temperature |
-40 °C to + 60 °C |
Battery |
9 volt Alkaline, NEDA 1604A/IEC 6LR61 |
Battery life |
400 hours typical, without backlight |
Measurement |
Range |
Resolution |
Accuracy ± ([% of reading] + [counts]) |
DC millivolts |
600.0 mV |
0.1 mV |
2.0 % + 3 |
DC volts |
6.000 V |
0.001 V |
|
60.00 V |
0.01 V |
600.0 V |
0.1 V |
Auto volts |
600.0 V |
0.1 V |
2.0 % + 3 (dc, 45 Hz to 500 Hz)
4.0 % + 3 (500 Hz to 1 kHz) |
AC millivolts1 true-rms |
600.0 mV |
0.1 mV |
1.0 % + 3 (dc, 45 Hz to 500 Hz)
2.0 % + 3 (500 Hz to 1 kHz) |
AC volts1 true-rms |
6.000 V |
0.001 V |
1.0 % + 3 (45 Hz to 500 Hz)
2.0 % + 3 (500 Hz to 1 kHz) |
60.00 V |
0.01 V |
600.0 V |
0.1 V |
Continuity |
600 ω |
1 ω |
Beeper on < 20 ω off > 250ω; detects opens or shorts of 500 µs or longer. |
Ohms |
600.0 ω |
0.1ω |
0.9 % + 2 |
6.000 kω |
0.001 kω |
0.9 % + 1 |
60.00 kω |
0.01 kω |
600.0 kω |
0.1 kω |
6.000 kω |
0.001 kω |
40.00 kω |
0.01 kω |
1.5 % + 2 |
Diode test |
2.000 V |
0.001 V |
0.9 % + 2 |
Capacitance |
1000 nF |
1 nF |
1.9 % + 2 |
10.00 µF |
0.01 µF |
100.0 µF |
0.1 µF |
9999 µF |
1 µF |
|
100 µF to 1000 µF |
|
1.9 % + 2 |
> 1000 µF |
|
5 % + 20 |
Lo-Z capacitance |
1 nF to 500 µF |
|
10 % + 2 typical |
>AC amps true-rms (45 Hz to 500 Hz) |
6.000 A |
0.001 A |
1.5 % + 3 |
10.00 A |
0.01 A |
>20 A continuous overload for 30 seconds max. |
>DC amps |
6.000 A |
0.001 A |
>1.0 % + 3 |
10.00 A |
0.01 A |
>20 A continuous overload for 30 seconds max. |
>Hz (V or A input)2 |
99.99 Hz |
0.01 Hz |
>0.1 % + 2 |
999.9 Hz |
0.1 Hz |
9.999 kHz |
0.001 kHz |
50.00 kHz |
0.01 kHz |
|